HeaderMain
5500sc Ammonia Monochloramine Analyzer
Category: Water Analysis
Sub Category: Process/Inline Analysis
Share
Add To Cart

5500sc Ammonia Monochloramine Analyzer

Applications:

  • Municipal Water
  • Tertiary Treatment

Feature

Details

Range

0.01 to 2 ppm (as N)

Parameter

Total Ammonia, Monochloramine, and Free Ammonia

Number of Sample Streams

1 or 2 in programmable sequence

Accuracy

± 5 % or 0.01 ppm (as N) for sample temperature 5 to 40 °C (41 to 104 °F); ±10% or 0.02 ppm for sample temperature 40 to 50 °C (104 to 122 °F), whichever is greater

Repeatability

3 % or 0.01 ppm (as N), whichever is greater

Lower Limit of Detection (LOD)

0.01 ppm (as N)

Response Time

Within 5 minutes

Reagent Consumption

1L of reagents every month with 5 minute cycle time

Operating Temperature Range

5 to 45 °C (41 to 113 °F)

Operating Humidity

5 to 95 % non-condensing

Sample Pressure

2.5 to 87 psi (0.17 to 6.0 bar)

Sample Temperature

5 to 50 °C (41 to 122 °F)

Sample Flow Rate

100 to 1000 mL/minute

Grab Sample

Grab Sample In and Grab Sample Out capability

Mounting

Wall, panel, or table

Inlet

6mm OD push-to-connect fitting for plastic tubing

Outlet

11mm (7/16 in.) ID slip-on fitting

Output

Four 0/4-20 mA Outputs, load impedance 600 ohm maximum

Communication

4-20mA Output, Optional: Modbus RS485, Profibus DP with external controller

Dimensions (H x W x D)

804 mm x 452 mm x 360 mm

Enclosure Rating

NEMA 4X/IP56

Certifications

CE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000)<br>cETLus (UL 61010-1:2012; NEMA 250:2003; CSA C22.2 No 61010-1:2012)<br>Australian RCM Marking

Weight

20.5 kg (45.2 lb) without reagents and standards, 30 kg (66.2 lb) with reagents, standards and cleaning solution

Warranty

1 year

Chloramination made easy.
The Hach 5500sc Ammonia Monochloramine Analyzer provides all the information you need to eliminate nitrification events and taste and odor issues, giving you total confidence in your process. Oh—and it’s easy, too.

Reliability without all the work
Only Hach can offer a low-maintenance solution with a pressurized reagent delivery system, easy-to-change reagent bottles, a simple troubleshooting menu, and PROGNOSYS predictive diagnostics. You’ll always know if your analyzer is doing its job, so you can do yours.

Easy operation
It has never been easier to monitor your chloramination process online. You’ll be up and running on day one with easy setup, and keep on running with at-a-glance status
LEDs, a user friendly interface, color-coded reagent bottles and a simple solution for verifying your results with grab samples.

Improved control of your chloramination process
Continuous online monitoring gives you a more accurate and complete picture of your chloramination process, allowing you to make data-driven decisions. With a dual
colorimeter design, you’ll get more results faster - keeping you one step ahead of nitrification.

5500sc Ammonia monochloramine.pdf
5500sc Ammonia monochloramine.pdf