5500sc Phosphate Low Range Analyzer
Category: Water Analysis
Sub Category: Process/Inline Analysis
Share
Add To Cart

5500sc Phosphate Low Range Analyzer

Applications:

  • Boiler Water
  • Steam Cycle
  • Water Cycle

Lower Maintenance, Less Downtime
Delivering reliable results that save you critical time and effort.

90 Days of Continuous Run Time
Only two liters of reagent are required for the analyzer to perform unattended for up to 90 days; twice as long as the Series 5000.

Save Time on Maintenance
The industry’s only pressurized reagent delivery system eliminates the frequent maintenance associated with pumps.

Avoid Downtime
Predictive diagnostic tools, including Hach’s proprietary Prognosys technology, warning LEDs, and high-visibility notification screens let you avoid unplanned downtime.

Clean, Fast and Easy Reagent Change
No more dripping reagents on the instrument, the floor, or your clothing while fumbling with tubes and straws. Simply match the color-coded cap to the sealed reagent bottle and twist gently.


Verify Easily with Hach Lab Products So You Don’t Waste Time Second-guessing

Grab Sample In and Grab Sample Out features allow quick analysis of a grab sample poured into the analyzer, and facilitate taking a sample out of the analyzer to verify in a lab test.

Feature

Details

Measuring Principle

Colorimetric

Range

LR model: 4-3000 µg/L as PO4 (Detection range capable of 4 - 5000 µg/L as PO4)

HR model: 200-50,000 µg/L as PO4

Accuracy

LR model: ±4 µg/L or ±4% of reading, whichever is greater

HR model: ±500 µg/L or ±5% of reading, whichever is greater

Repeatability

LR model: ±1%

HR model: ±500 µg/L or ±5% of reading, whichever is greater

Lower Limit of Detection (LOD)

LR model: 4 µg/L

HR model: 200 µg/L

Response Time

Typically, 9.5 minutes at 25 °C (77 °F); changes with temperature

Reagent Consumption

2 L of each reagent every 90 days with 15-minute cycle time

Operating Temperature Range

5 - 45 °C (41 - 113 °F)

Operating Humidity

5 - 95 % RH non-condensing (indoor use only)

Sample Pressure

0.14 - 6 bar (2 - 87 psi) (to Preset Pressure Regulator)

Sample Temperature

5 - 50 °C (41 - 122 °F)

Sample Flow Rate

55 - 300 mL/min

Number of Channels

1, 2, 4; programmable sequence

Grab Sample

Grab Sample In and Grab Sample Out capability

Mounting

Wall, panel, or table

Fitting

Sample line and sample bypass drain: 6 mm (¼-in.)

Air purge inlet: 6 mm (¼-in.)

Power Requirements (Voltage)

100 - 240 V AC, 24 V DC

Power Requirements (Hz)

50/60 Hz

Output

4 - 20 mA

Dimensions (H x W x D)

804 mm x 452 mm x 360 mm

Enclosure Rating

NEMA 4X/IP56

Certifications

CE: EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000

KC: EN 61326-1: 2006

C-tick: EN 61326-1: 2006

cETLus: UL 61010-1: 2012; NEMA 250: 2003; CSA C22.2 No 61010-1: 2012

5500sc Phosphate analyzer.pdf
5500sc Phosphate analyzer.pdf