Lower Maintenance, Less Downtime
Delivering reliable results that save you critical time and effort.
90 Days of Continuous Run Time
Only two liters of reagent are required for the analyzer to perform unattended for up to 90 days; twice as long as the Series 5000.
Save Time on Maintenance
The industry’s only pressurized reagent delivery system eliminates the frequent maintenance associated with pumps.
Avoid Downtime
Predictive diagnostic tools, including Hach’s proprietary Prognosys technology, warning LEDs, and high-visibility notification screens let you avoid unplanned downtime.
Clean, Fast and Easy Reagent Change
No more dripping reagents on the instrument, the floor, or your clothing while fumbling with tubes and straws. Simply match the color-coded cap to the sealed reagent bottle and twist gently.
Verify Easily with Hach Lab Products So You Don’t Waste Time Second-guessing
Grab Sample In and Grab Sample Out features allow quick analysis of a grab sample poured into the analyzer, and facilitate taking a sample out of the analyzer to verify in a lab test.
Feature | Details |
Measuring Principle | Colorimetric |
Range | LR model: 4-3000 µg/L as PO4 (Detection range capable of 4 - 5000 µg/L as PO4) HR model: 200-50,000 µg/L as PO4 |
Accuracy | LR model: ±4 µg/L or ±4% of reading, whichever is greater HR model: ±500 µg/L or ±5% of reading, whichever is greater |
Repeatability | LR model: ±1% HR model: ±500 µg/L or ±5% of reading, whichever is greater |
Lower Limit of Detection (LOD) | LR model: 4 µg/L HR model: 200 µg/L |
Response Time | Typically, 9.5 minutes at 25 °C (77 °F); changes with temperature |
Reagent Consumption | 2 L of each reagent every 90 days with 15-minute cycle time |
Operating Temperature Range | 5 - 45 °C (41 - 113 °F) |
Operating Humidity | 5 - 95 % RH non-condensing (indoor use only) |
Sample Pressure | 0.14 - 6 bar (2 - 87 psi) (to Preset Pressure Regulator) |
Sample Temperature | 5 - 50 °C (41 - 122 °F) |
Sample Flow Rate | 55 - 300 mL/min |
Number of Channels | 1, 2, 4; programmable sequence |
Grab Sample | Grab Sample In and Grab Sample Out capability |
Mounting | Wall, panel, or table |
Fitting | Sample line and sample bypass drain: 6 mm (¼-in.) Air purge inlet: 6 mm (¼-in.) |
Power Requirements (Voltage) | 100 - 240 V AC, 24 V DC |
Power Requirements (Hz) | 50/60 Hz |
Output | 4 - 20 mA |
Dimensions (H x W x D) | 804 mm x 452 mm x 360 mm |
Enclosure Rating | NEMA 4X/IP56 |
Certifications | CE: EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000 KC: EN 61326-1: 2006 C-tick: EN 61326-1: 2006 cETLus: UL 61010-1: 2012; NEMA 250: 2003; CSA C22.2 No 61010-1: 2012 |