Range | 0 - 10 mg/L free or total residual chlorine |
Accuracy | ± 5% or ± 0.04 mg/L (whichever is greater) from 0 - 5 mg/L Cl2 ± 10% from 5 - 10 mg/L Cl2 |
Lower Limit of Detection (LOD) | 0.03 mg/L |
Cycle Time | 2.5 minutes |
Inlet Pressure | 0.3 - 5.2 bar (4.5 - 75 psig) supplied to Y-strainer 0.1 - 0.3 bar (1.5 - 5 psig) supplied to analyzer |
Inlet | ¼-inch OD polyethylene tube, quick-disconnect fitting |
Drain | ½-inch ID flexible hose |
Air Purge | Optional with 3/8-inch quick-connect fitting and tubing; 0.003 m3/min at 1.38 bar (20 psig) maximum |
Sample Requirements | Y-strainer filtration with 40-mesh screen or higher |
Sample Flow Rate | 60 - 200 mL/min through the instrument |
Sample Temperature | 5 - 40 °C (41 - 104 °F) |
Operating Temperature Range | 5 - 40 °C (41 - 104 °F) |
Operating Humidity | 0 - 90% relative humidity, non-condensing |
Interferences | Other oxidizing agents such as bromide, chlorine dioxide, permanganate, and ozone will cause a positive interference. Hardness must not exceed 1,000 mg/L CaCO3. |
Reagent Consumption | 0.5 L of buffer solution and 0.5 L of indicator solution in 31 days |
Certifications | CE compliant with: EN 61326-1, CISPR 11, EN 50581 ACMA RCM South Korea KC Certificate North America: FCC Supplier’s Declaration of Conformance, IEC/EN 60529, ICES-003 |
Light Source | LED, measurement at 510 nm, 1 cm light pathlength |
Enclosure Rating | IP66 |
Mounting | Wall mount |
Power Requirements (Voltage) | 12 VDC, 400 mA maximum (supplied by the controller) |
Dimensions (H x W x D) | 342 mm x 329 mm x 177 mm |
Weight | 4.1 kg (9.0 lbs.) |
The Next Standard in Chlorine Analysis
Built on a legacy of reliability in online chlorine analysis, the advanced design of the CL17sc reduces routine maintenance touch time and provides powerful diagnostic features and enhanced connectivity. The result is less hassle, minimized risk of data loss, and even more reliable information to empower your decision making.
Maintenance made easy
The CL17sc reduces your routine maintenance touch time with programmable alerts, simplified tubing replacement, and step-by-step maintenance instructions.
Peace of mind through comprehensive diagnostics
With upgraded features like a flow meter, colorimeter window, multi-color status light, and predictive diagnostic software, you know your instrument is operating as intended.
Expanded connectivity. Increased flexibility.
By pairing the CL17sc with Hach’s SC controller platform, your options increase significantly: internal data logging; external analog and digital communication alternatives; and multiparameter instrument flexibility.
The CL17sc is compliant with US EPA regulation 40 CFR 141.74. Both Method 4500-CL G and Method 334.0 can be used for measuring residual chlorine in drinking water.