The smart choice for accurate and reliable online process pH/ORP measurement
Hach Digital pHD sc sensors are available in convertible (PEEK or PPS), insertion, and sanitary body styles. Three electrodes are used in these
sensors to increase measurement accuracy and eliminate sensor ground loops.

Exceptional Process Sensor Performance with the Differential Electrode pHD Measurement Technique
This field-proven technique uses three electrodes instead of the two normally used in conventional pH/ORP sensors. Process and reference electrodes measure the pH/ORP differentially with respect to a third ground electrode. The end result is unsurpassed measurement accuracy, reduced reference junction potential, and elimination of sensor ground loops. These process pH/ORP sensors provide greater reliability, resulting in less downtime and maintenance.

Lower Maintenance Needs with the Double Junction Salt Bridge
The double junction salt bridge creates a barrier to contamination which minimizes the dilution of the internal standard cell solution. The result is lower maintenance needs and a longer time period between calibrations.

Extended Working Life with the Replaceable Salt Bridge/Protector
The unique, replaceable salt bridge holds an extraordinary volume of buffer to extend the working life of the sensor by protecting the reference electrode from harsh process conditions. The salt bridge simply threads onto the end of the sensor if replacement is needed.

Reliability with Built-in Encapsulated Preamp
Encapsulated construction protects the sensor’s built-in preamp from moisture and humidity, ensuring reliable sensor operation. The preamp in the pHD analog sensor produces a strong signal, enabling the sensor to be located up to 1000 m (3280 ft.) from the analyzer.

Innovative Technology
The former GLI, now a Hach Company brand, invented the Differential Electrode Technique for pH measurement in 1970. The pHD sensor series takes this field-proven technology to a new level.

Versatile Mounting Styles
Sensors are available in four mounting styles - convertible, insertion, immersion, and sanitary

Differential Sensor Warranty
Hach Company offers the best sensor warranty in the industry on its Differential Sensors. We will replace any Differential Sensor that fails due to defects in materials or workmanship within one year from the date of shipment-and up to 30 months on a prorated basis for any failure. 

 

Compliance

For hazardous and maritime applications, CE

For non-hazardous and non-maritime applications only

Operating Temperature Range

-5 - 70 °C (23 - 158 °F) pHD and ORP

0 - 50 °C (32 - 122 °F) SS pHD

Before initial pH calibration, calibrate the temperature measurement when the sensor is in water or buffer which is at approximately the same temperature as the pH buffers (matches current recommendation)

-5 - 70 °C (23 - 158 °F) pHD and ORP

0 - 50 °C (32 - 122 °F) SS pHD

Before initial pH calibration, calibrate the temperature measurement when the sensor is in water or buffer which is at approximately the same temperature as the pH buffers (matches current recommendation) Please Note: When the sensor is placed into the application sample, if that sample is more than 10 °C (18 °F) different than the previous temperature/pH calibration, then it is recommended to recalibrate the temperature while the sensor is in the sample to maintain the ±0.5 °C (±0.9 °F) temperature accuracy specification.

Range

-2.0 to 14.0 pH -1500 to +1500 mV ORP

Sensitivity

± 0.01 pH ± 0.5 mV

Flow Rate

3 m (10 ft.) per second, maximum

Pressure Range

Digital: 6.9 bar at 70 ºC (100 psi at 158 ºF)

Temperature Sensor

NTC 300 Ω thermistor for automatic temperature compensation and analyzer temperature readout

Transmission Distance

100 m (328 ft.), maximum

Temperature Accuracy

± 0.5 °C (± 0.9 °F)

Sensor Cable

4 conductor cable with one shield and polyurethane jacket; rated to 105 °C (221 ºF); 10 m (33 ft.) standard length

Wetted Materials

PEEK or PPS, salt bridge of matching material with PVDF junction, glass process electrode, titanium ground electrode, and FKM/FPM O-ring seals (pH sensor with optional HF-resistant glass process electrode has 316 stainless steel ground electrode, and perfluoroelastomer wetted O-rings; consult factory for other available wetted O-ring materials)