HeaderMain
Filtrax sample filtration system
Category: Water Analysis
Sub Category: Process/Inline Analysis
Share
Add To Cart

Filtrax sample filtration system

Applications:

  • Filtration
  • Industrial Wastewater
  • Municipal Water
  • Nutrient Monitoring
  • Primary Treatment
  • Secondary Treatment
  • Tertiary Treatment

Established Technology In Situ and Down the Line
The Hach Filtrax Sample Filtration System is specifically designed for efficient wastewater analysis. Samples are extracted directly from the activated sludge aeration basin or final settling tank and solids are removed in situ. Separate, high-maintenance submersible pumps become unnecessary. Sample delivery lines and analyzers stay clean. Sample extraction automatically alternates between two continuously cleaned filter elements to allow optimum cleaning of the idle filter. Sample pre-treatment and delivery is continuous and efficient—even with a high sludge volume index or floating sludge.

Simple to Use and Maintain
A built-in air purge system in the Filtrax system prevents adhesion of solids to the filter membranes and greatly reduces maintenance cost and time. All tubing is completely accessible and easy to replace. Moving parts never come into contact with the sample. Self-diagnostics provide operators with three levels of error messages if components require attention. Operators control all Filtrax system functions with menu-driven software. The instrument display provides continuous readout of current flow rate through each filter module as well as heater status. Two programmable alarm relays are provided—one relay can be used to alert operators to inspect the Filtrax system when flow decreases and a second programmable relay can shut down the unit if flow decreases further.

Operating Principle
All control components are heated and contained in an IP55-rated enclosure, suitable for outdoor installation. The Filtrax system is made of three components.
Control unit: The control unit is internally heated for outdoor installation. Inside are sample pumps, an air compressor for filter membrane cleaning, and the electronic controller. Filters are constantly cleaned by a vigorous stream of air bubbles against the sides of the module. Two small peristaltic pumps alternately draw wastewater sample through the two filter membranes for optimum cleaning of the idle filter.
Filtration module holder: The filtration module holder is made of stainless steel and is designed for submerged installation. It contains two filter modules each equipped with an ultra-filtration (0.15 µm) filter membrane.
Sample delivery hoses: Samples are pulled to the control unit from the filtration module holder with a 15-ft. (5-m) heated suction hose. This hose also contains the air compressor tubing. Samples are then delivered directly to the on-line measuring instrument.

Hach’s Family of Nutrient Analyzers
The Hach Filtrax Sample Filtration System is designed to optimize performance of Hach’s Amtax™ sc Ammonia Analyzer or Phosphax™ sc Phosphate Analyzer. It also
complements Hach’s NITRATAX plus sc or NITRATAX clear sc Nitrate Analyzer by supplying solids-free sample to its flow through panel. Rely on the Filtrax system to provide properly conditioned sample to Hach’s APA 6000™ Alkalinity, Ammonia, Monochloramine, and Phosphate Analyzers too.

Feature

Details

Sample Flow

Approx. 900 mL/h for up to three process instruments

Sample Temperature

41 to 104°F (5 to 40°C)

Ambient Temperature

-4 to 104°F (-20 to 40°C)

Suction Head Space

9 ft. (3 m) maximum (filter module holder to controller)

Maximum Sample Delivery Head Space

21 ft. (7 m) maximum (control unit to analyzer)

Cable Length

Suction Hose: 5 m (heated)

Sample Delivery Hose: 2 m (heated or unheated), 10 m (heated), 20 m (heated), or 30 m (heated)

Outputs

Programmable Fault Alarm Contact: Potential free contact (230 V, 3A maximum)

Programmable Warning Contact: Potential free contact (230 V, 3A maximum)

Service Interface: RS232

Power Supply

115 or 230 Vac ±10% AC, 50/60 Hz

Inspection Requirement

Approximately 1 hour/month

Enclosure Class

IP55 (outdoor installation)

Certification

CE, UL, CSA

Weight (approximate)

Control Unit: 49 lb. (22 kg)

Module Holder with 5-m Suction Tube: 20 lb. (9 kg)

Sample Tube, 33-ft. (5-m): 11 lb. (5 kg)

Sample Tube, 66-ft. (20-m): 22 lb. (10 kg)

Mounting pipe, 6.6-ft. (2-m): 11 lb. (5 kg)

Dimensions (approximate)

Control Unit: 17 x 21 x 9 in. (430 x 530 x 220 mm)

Module Holder: 4 x 20 x 13 in. (92 x 500 x 340 mm)

Filtrax.pdf