Category: Fuels and Oil Analysis
Sub Category: Viscosity/Distillation/Flash Point
Share
Add To Cart

Flash point on petroleum products having a flash point between -18°C and 71°C (kerosene and solvents). Suitable for flash point detection on different substances and waste materials, solvents…

Measuring Abel Principle

The sample is warmed up according to the methods. When the sample reaches the selected test temperature, the shutter is opened and the ignition system introduces itself automatically. If the flash point is reached, the detection is done by an ionisation detector. If not, the shutter closes again and the sample continues to warm up until the next test temperature.

EN 924 - EN 13736
IP 170 - IP 491 - IP 492
ISO 1516 - ISO 3679 - ISO 13736

kerosene and solvents

Measuring Abel Devices
  • Measurement of the Flash Point detected by an ionisation detector
  • Testing unit equipped with two ignition systems
  • Electrical system or flame exposure device
Measuring Temperature Probe
  • Platinum resistance PT100 class A
Measuring Parameters
  • Temperatures: in °C
  • Measuring range: -50°C … +120°C
  • Resolution: 0.06 °C
  • Accuracy: ± 0.1 °C
  • Repeatability / Reproducibility: as per standards methods or better
Electrical Supply
  • 220V ± 15% / 50 to 60 Hz
  • 115V ± 15% / 60 Hz
Cord Cable
  • 3 conductors flexible cable with schuko plug
Ambient Temperature
  • Max 35°C
  • H.R. 80%

Dimensions
  • width 37 cm
  • depth 48 cm
  • height 61 cm
Weight
  • 32 kg