HeaderMain
Orbisphere M1100 Luminescent Dissolved Oxygen Sensor
Category: Water Analysis
Sub Category: Process/Inline Analysis
Share
Add To Cart

Orbisphere M1100 Luminescent Dissolved Oxygen Sensor

M1100 (Low Level Sensor)

Range

0 - 2000 ppb DO (dissolved oxygen)

Temperature Range

Accurate from -5 - 50 °C (23 - 122 °F)

Resistant to temperature from -5 - 100 °C (23 - 212 °F)

Repeatability

± 0.4 ppb or 1 % whichever is greater

Reproducibility

± 0.8 ppb or 2 % whichever is greater

Accuracy

± 0.8 ppb or 2 % whichever is greater

Lower Limit of Detection (LOD)

0.6 ppb

Response Time

(90%)< 10 s (gas phase)

Display Resolution

0.1 ppb

Calibration

Single point zero calibration with standard 99.999% nitrogen (quality 50) or equivalent oxygen free gas

Sample Pressure

1 - 20 bar absolute

 

 

M1100 (High Level Sensor)

 

 

Range

0 - 40 ppm DO (dissolved oxygen)

Temperature Range

Accurate from -5 - 50 °C (23 - 122 °F)

Resistant to temperature from -5 - 100 °C (23 - 212 °F)

Repeatability

± 0.015 ppm or 2 % whichever is greater

Reproducibility

± 0.02 ppm or 3 % whichever is greater

Accuracy

± 0.02 ppm or 3 % whichever is greater

Lower Limit of Detection (LOD)

0.015 ppm

Response Time

(90%)< 10 s (gas phase)

Display Resolution

0.1 ppb

Calibration

Two points at cap replacement (zero and air), one during use (air)

Sample Pressure

1 - 20 bar absolute

 

 

Orbisphere Controller

 

 

Enclosure Construction

Wall (pipe) mount: stainless steel

Panel mount: aluminum

Enclosure Rating

Wall (pipe) mount: IP65, NEMA 4X

Panel mount: IP65

Compliance

EMC: EN61326-1:2006

Certifications

CE: EN61010-1:2010

ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1

Display

Colour TFT touch-screen display

Analog Outputs

3 smart 0/4-20 mA (500 ohms), programmable as linear or tri-liniear, configurable to send diagnostics information

Relays

3 measurement alarm relays (2A to 30 VAC or 0.5 A to 50 VDC); configurable to send diagnostics information

 

1 system alarm relay (2 A to 30 VAC or 0.5 A to 50 VDC)

Communication

RS485

Profibus DP (Optional)

Ethernet

USB-client to download data from a computer

USB-host to download data with a USB memory stick

Data Storage

Rolling buffer or store once mode for up to 1000 measurements and 1000 operator actions

Holds calibration records for last 50 calibrations

User Interface

Touch-screen panel displays: concentration, trend graph, diagnostics, alarm status, historical data

Dimensions (H x W x D)

230.5 mm x 250 mm x 160 mm

Power Supply

Universal 100 - 240 VAC @ 50/60 Hz, 25 VA

10 - 36 VDC, 25 W

Monitors oxygen in the beverage production process
The Orbisphere M1100 optical sensor, together with the Orbisphere 410 one channel and the Orbisphere 510 multichannel controller, offers a new way of monitoring oxygen in the beverage production process. 

 

Minimal Drift and Annual Calibration
The M1100 sensor provides immediate oxygen readings with a measurement frequency of two seconds. The instrument carries consistent readings with no drift for 12 months without calibration (when utilising standard weekly CIP processes), surpassing other optical sensors that display significant drift after only a few months in similar conditions.


Minimal Maintenance Optical Technology
The M1100 pioneered the use of luminescent technology in brewing applications and as a result, the sensor doesn’t require the replacement of membranes or any electrolytes. Additionally, the sensor’s accuracy is unaffected by process changes or pressure shocks, further reducing maintenance.
Annual maintenance is limited to just a few minutes for a zero-point calibration. Chemicals are not required for this process, making the task easier and safer without reducing measurement precision.


Low Level Oxygen Measurement with Accurate ppb
The M1100-L sensor has a lower detection level of 0.6 ppb. This highly accurate instrument’s readings are essential to control low oxygen levels in beverage production. As an example, it is critical to control low level oxygen in the post fermentation applications of the brewery to ensure high quality of the final product.

High Level Oxygen Measurement with Accurate ppm
The M1100-H sensor has a range of 0-40 ppm, and is ideal for use in wort applications, for example. Even in this harsh wort environment, the instrument maintains very good accuracy and minimal drift. Only a yearly maintenance and calibration will be required in most cases.