HeaderMain
Phosphax sc Phosphate Analyzer
Category: Water Analysis
Sub Category: Process/Inline Analysis
Share
Add To Cart

Phosphax sc Phosphate Analyzer

Applications:

  • Industrial Wastewater
  • Municipal Water
  • Nutrient Monitoring
  • Secondary Treatment
  • Tertiary Treatment

The Phosphax sc online analyzer provides reliable and accurate PO4 measurements

Multiple measurement ranges for a variety of wastewater applications
With detection limits as low as 0.05 mg/L and as high as 50 mg/L, the Phosphax sc phosphate analyzer can be used anywhere in the wastewater treatment process, from
the influent or start of the phosphorus removal process where phosphate levels may be high to the effluent where phosphate levels are at their lowest.


Low cost of operation with proven yellow method
The Phosphax sc analyzer determines ortho-phosphate concentration using the molybdovanadate yellow colorimetric method which optimizes reagent consumption and helps save on operating costs.

Generate actionable insights from measurement data
The Phosphax sc is Claros enabled so you can leverage the Hach Water Intelligence System to collect, manage and analyse data from your instrument. 

Easy installation at the measurement point
Hach’s Phosphax sc phosphate analyzer is designed to be installed at the measurement point (indoor and outdoor options). The housing is weatherproof and lockable for
installation at the basin, even in the toughest climates. Mounting options include: wall, rail, or standing. The unit comes complete and assembled; no separate housing is
required.


Low maintenance
Several features make the Phosphax sc phosphate analyzer easy to use and maintain:
1. Automatic cleaning at customized intervals.
2. Automatic zero-calibration at each measuring cycle.
3. Prognosys Predictive Diagnostics alerts you to upcoming instrument issues and guides you on whether the changes in your measurements are due to your instrument or your water.
4. Easy access to reagents and wear parts.

Feature

Details

Range

Low Range: 0.05 - 15.0 mg/L PO4-P

High Range: 1 - 50 mg/L PO4-P

Lower Limit of Detection (LOD)

Low Range: 0.05 mg/L PO4-P

High Range: 1.0 mg/L PO4-P

Accuracy

Low Range: 2% ± 0.05 mg/L

High Range: 2% ± 1.0 mg/L

Reproducibility

Low Range: 2% + 0.05 mg/L

High Range: 2% + 1.0 mg/L

Reagent Consumption

Low Range: 500 mL/month

High Range: 1000 mL/month

Response Time

< 5 minutes

Measurement Method

Photometric method using vanadate-molybdate

Measuring Interval

5 - 120 min, adjustable per 5 min

pH Range

5 - 9 pH

Pressure Range

-30 - 50 mbar with continuous sample preparation; at overflow vessel

Permissible Chloride Range

Max. Cl- concentration: 1000 mg/L

Operating Conditions

Indoor model: 5 - 40 °C; 95% relative humidity, non-condensing

Outdoor model: -20 - 45 °C; 95% relative humidity, non-condensing

Sample Temperature

4 - 40 °C

Sample Quality

Ultra filtrated or comparable

Flow

1 - 20 L/h sample (free of suspended solids)

Power Requirements (Voltage)

115 - 230 VAC, 50/60 Hz, power provided by SC controller or power box

Dimensions (H x W x D)

Indoor model: 720 mm x 540 mm x 370 mm

Outdoor model: 720 mm x 540 mm x 390 mm

Cable Length

2 m fixed data cable at analyzer

Weight

Indoor model: 29 kg (without sample preparation system and chemicals)

Outdoor model: 31 kg

Material

ASA/PC UV-resistant

Enclosure Rating

Indoor model: IP54

Outdoor model: IP55

Phosphax sc.pdf