HeaderMain
TSS sc Turbidity and Suspended Solids Inline Probe
Category: Water Analysis
Sub Category: Process/Inline Analysis
Share
Add To Cart

Turbidity and suspended solids probe for use in Triclamp piping systems

The TSS XL sc is a digital process probe for the extremely accurate in pipe measurement of turbidity and suspended solids specially designed for use in beverage industries. This in pipe probe is installed by XL weld-on fitting and can measure both on-line suspended solids and turbidity in one instrument. The two channel 90° scattered light turbidity method is in accordance with DIN EN ISO 7027. For the determination of suspended solids a scattered light measurement method at 120 ° is integrated. This eight channel measurement system, with an integrated bubble and temperature compensating software, enables the instrument to have a wide measuring range. The TSS XL sc probes are compatible with the Hach SC200 and SC1000 controller platform. Furthermore, they are made of polished stainless steel with a scratch resistant and easy to clean sapphire window design to withstand harsh environments.

Benefits

Measures online suspended solids in virtually all applications under the strictest regulatory conditions
TSS sc probes measure online suspended solids from spring water to corrosive chemicals under the strictest regulatory conditions. With 8 different probe models with several mounting options, it is possible to measure suspended solids over a wide range of concentrations with just one probe.

Unique Multi-Beam Pulsed Infrared Light System
TSS sc probes have a double optical system with two pulsating infrared LEDs and four receivers. This combined multiple beam alternating light method with beam focusing facilitates accurate color independent measurement of turbidity from 0.001 to 4,000 FNU and suspended solids from 0.001 to 500 g/L.

Automatic Compensation for Air Bubbles and Temperature Swings for Accuracy
Special software gives the system the ability to recognize gas bubbles or temperature swings for a more accurate suspended solids or turbidity measurement.

Standard-Compliant and Precise with Easy Calibration
The turbidity measurement complies with the standard DIN EN 27027 (ISO 7027). Calibration is not necessary. If the probe is used to measure the solids content, a one-point calibration suffices. For special applications, curves can be defined using several calibration points.

Withstands Difficult Conditions and Harsh Environments
Probes are made of highly polished stainless steel or titanium with a scratch resistant sapphire window. They are designed to withstand harsh environments and keep particles from sticking to the surface. 

Models

TSS sc

TSS W sc

TSS HT sc

TSS Vari sc

TSS XL sc

TSS EX1 sc

TSS Titanium2 sc

TSS Titanium7 sc

Mounting

Tank-side, Triclamp, Inline, Varivent

Configurations

measuring tube, XL measuring tube

Measurement

Method

Combined multiple beam alternating light method with infrared diode system and beam focusing

Turbidity (TRB): 2-channel

90º scattered light measurement in accordance with DIN EN ISO 7027,

wavelength = 860 nm

Solids (TSS): 90º and 120º scattered light measurement,

wavelength = 860 nm

EExp / Hazardous Location

Class I Div 2 certified (TSS EX1 sc only)

Range

Turbidity (TRB): 0.001 - 9999 NTU

Solids (TSS): 0.001 - 500 g/L

Accuracy

Turbidity: Up to 1000 FNU/NTU:

< 5 % of measurement value or

±0.01 NTU, whichever is greater

Repeatability

Turbidity (TRB): < 3%

Solids (TSS): < 4%

Response Time

1 s < T90 < 300 s (adjustable)

Calibration

Turbidity (TRB): Factory calibrated

Solids (TS): To be calibrated

by customer on site

Zero point: Permanently calibrated

in the factory

Pressure Range

< 6 bar: TSS W sc

< 10 bar: TSS sc, TSS HT sc,

TSS Titanium2 sc, TSS Titanium7 sc

≤ 10 bar: TSS EX1 sc

< 16 bar: TSS Vari sc, TSS XL sc

Ambient Temperature

0 - 50 °C: TSS W sc

0 - 60 °C: TSS sc, TSS Titanium2 sc,

TSS Titanium7 sc

0 - 80 °C: TSS Vari sc, TSS XL sc

0 - 90 °C: TSS HT sc

-10 - 50 °C: TSS EX1 sc

Dimensions

Tank-side models, installation sensor

Inline models:

330 mm x 40 mm (L x D)

Installation sensor Triclamp models:

332 mm x 40 mm (L x D)

TSS Vari sc, TSS XL sc:

232 mm x 40 mm (L x D)

Weight

Tank-side models, installation sensor

Inline/Triclamp models: approx. 1.6 kg

TSS Vari sc, TSS XL sc: approx. 1.5 kg

TSS EX1 sc: approx. 2.7 kg

Controller

Compatibility

SC200, SC1000

TSS sc Suspended Solids Probes.pdf
TSS sc Suspended Solids Probes.pdf