Models:
TSS sc
TSS W sc
TSS HT sc
TSS Vari sc
TSS XL sc
TSS EX1 sc
TSS Titanium2 sc
TSS Titanium7 sc
Turbidity and suspended solids probe for use in VARIVENT® piping systems
The TSS Vari sc is a digital process probe for the extremely accurate in pipe measurement of turbidity and suspended solids specially designed for use in food and pharmaceutical sectors. This in pipe probe is installed by Varivent access unit and can measure both on-line suspended solids and turbidity in one instrument. The two channel 90° scattered light turbidity method is in accordance with DIN EN ISO 7027. For the determination of suspended solids a scattered light measurement method at 120 ° is integrated. This eight channel measurement system, with an integrated bubble and temperature compensating software, enables the instrument to have a wide measuring range. The TSS Vari sc probes are compatible with the Hach SC200 and SC1000 controller platform. Furthermore, they are made of polished stainless steel with a scratch resistant and easy to clean sapphire window design to withstand harsh environments.
TSS sc probes have a double optical system with two pulsating infrared LEDs and four receivers. This combined multiple beam alternating light method with beam focusing facilitates accurate color independent measurement of turbidity and suspended solids.
Special software enables the system to recognize gas bubbles or temperature swings for a more accurate suspended solids or turbidity measurement.
The turbidity measurement complies with the standard DIN EN 27027 (ISO 7027). Calibration is not necessary. If the probe is used to measure the solids content, a one-point calibration suffices. For special applications, curves can be defined using several calibration points.
Measurement data is displayed and processed with the help of a digital Hach SC200 or SC1000 controller.
Specification | Technical Attributes |
Accuracy | Turbidity up to 1000 NTU: < 5% of the measured value ±0.01 NTU |
Ambient Temperature | 0 - 60 °C (briefly 80 °C) |
Application | Varivent piping systems |
Automatic Wiper | No |
Cable Length | 10 m (optional extension cables available) |
Calibration | Turbidity (TRB): Factory calibrated |
Calibration Method | Turbidity (TRB): Factory calibrated |
Controller Compatibility | SC200, SC1000, SC4500. All controllers sold separately |
Diameter | 40 mm |
Flow | Max. 3 m/s (the presence of air bubbles affects the measurement) |
Includes | Turbidity & Suspended Solids sensor, user manual |
Installation Style | Varivent Measuring tube |
Length | TSS VARI sc 232 mm |
Maintenance Interval | 1 h/month |
Material | Gasket: FKM; Wiper: PA (GF), TPV |
Max Temperature | 80 °C |
Measurement Method | Combined multiple beam alternating light method with infrared diode system and beam focusing |
Measuring Principle | Combined multiple beam alternating light method with infrared diode system and beam focusing |
Model | TSS Vari sc |
Mounting Configurations | VARIVENT |
Operating Temperature Range | 0 - 80 °C |
Parameter | Turbidity, Suspended Solids |
Pressure Range | <16 bar or <160 m |
Range | Turbidity (TRB): 0.001 - 9999 NTU |
Repeatability | TSS content: < 4 % |
Response Time | 1 s < T90 < 300 s (adjustable) |
Response Time T90 | 1 - 300 s adjustable |
Special Notes | Installation note: |
Weight | Approx. 1.5 kg |
What's Included? | Turbidity & Suspended Solids sensor, user manual |