HeaderMain
XPLORER
Category: Fuels and Oil Analysis
Sub Category: Combustion analyzers (Halogens, Sulfur, N2…)
Share
Add To Cart

TE Instruments has developed the XPLORER, an AOX, TOX, POX and EOX analyzer offering fast, accurate and precise analysis of organic halogens.

This brand new model is designed to offer customized solutions to match both current and future analytical needs.

  • Compact design, smallest elemental combustion analyzer available in today’s market
  • Fast generation of sample queues and application methods with TE Instruments Software (TEIS)
  • Short start-up time (less than 15 minutes)
  • Fast and precise measurement of soil and liquid samples
  • Easy to use and intuitive user interface
  • Compact, stackable auto sampler for high sample throughput and low cost per analysis
  • Ultra-low detection limit, high stability and reliability due to the temperature-controlled titration cell
  • Low maintenance, optimal combustion and conditioning of gases results in near to zero downtime
  • Fast and easy switching between AOX, TOX, POX and EOX analysis, resulting in high productivity

AOX/TOX

EOX

POX

DIN 38409-part 14

DIN 38409-part 8

NEN 6401

DIN 38414-part 18

NEN 5777

 

ISO 9562

NEN 5735

 

ISO 11480

NEN 6402

 

EPA 9020

NEN 6676

 

EN 1485

NEN 6401

 

Fast screening of organic halogens greatly relies on the analysis of AOX, TOX, POX and EOX content. TE Instruments coulometric systems are the new bench mark for many applications:

  • Drinking water
  • Surface water
  • Ground water
  • Effluent water
  • Influent water
  • Waste water
  • Cooling water
  • Salty water
  • Process water
  • Pulp and paper products
  • Soil
  • Sediment
  • Sludge and waste oil

Dimensions (W x H x D)

36 x 27.2 x 69 cm (14.2 x 10.7 x 27.2 inch)

Weight

29 kg (64 lbs)

Voltage

100-240 V, 50-60 Hz

Power requirement (max)

1150 W

Gas connectors

⅛" Swagelok

Gases

Oxygen 99.6 % (2.6), Argon 99.998 % (4.8)

Input gas pressure

3-10 bar

Internal gas pressure

1.8 bar, adjustable

Furnace voltage

Dual zone, low voltage

Furnace temp. (max)

1150 °C (2100 °F)

 Furnace cooling

Pulling Fan, auto control

Sample introduction AOX/TOX

Solid introduction by quartz boat

Sample introduction EOX

Direct injection by liquids module

Sample introduction POX

Introduction by POX module

Solids

5-1000 mg

Boat driver

Software controlled, adjustable

Slider/shutter driver

Software controlled, adjustable

Detector

SMD, Digital Coulometer

Detector accuracy

Better than 2% CV

Titration cell conditioning

Temperature controlled, adjustable

Software

dot.NET-based, TEIS software

Ambient temperature

5-35 °C (41-95 °F)

Xplorer-MCT.pdf