Clora®Total Chlorine Analysis


RPV VALVES

Clora benchtops are compliant with ASTM D7536 and D4929 standards, and deliver a limit of detection of 0.13 ppm and a dynamic range up to 3000 ppm. Clora is powered by MWDXRF, the same technology found in our signature sulfur analyzer, Sindie. This direct measurement technique does not require gasses or high temperature processes equating to easy operation and minimum maintenance requirements.

Clora offers many features such as extended range with and without catalyst, autosampler, and Accu-flow.

Accu-flow technology eliminates particle settling and delivers results that better reflect crude streams as they exist in the refinery. Over a typical measurement cycle, the heavier particles can settle to the bottom of the sample cup and cause higher than normal results.

Key advantages

  • LOD: 0.13 ppm for hydrocarbons
  • LOD: 0.3 ppm for aqueous samples
  • Dynamic Range: 0.13 ppm to 3000 ppm
  • Fits on any lab bench
  • Easy to use:
  1. Intuitive touch screen 
  2. Just plug-in and measure
  3. Measurement time: 30-900 s
  • Extremely low maintenance: no conversion gasses, heating elements, columns, or quartz tubing

TRUSTED PRECISION

Monochromatic Wavelength Dispersive X-ray Fluorescence (MWDXRF®) utilizes state-of-the-art focusing and monochromate optics to increase excitation intensity and dramatically improve signal-to-background over high power traditional WDXRF instruments. This enables significantly improved detection limits and precision, and a reduced sensitivity to matrix effects. A monochromatic and focused primary beam excites the sample and secondary characteristic fluorescence X-rays are emitted from the sample. A second monochromating optic selects the chlorine characteristic X-rays and directs these X-rays to the detector. MWDXRF is a direct measurement technique and does not require consumable gasses or sample conversion

Specifications

Model

Clora

Calibration

8 calibration curves. Automatic and manual calibration functionality

Test Method

ASTM D7536 and D4929

Sample Cup Volume

Sample Cup Volume 10 ml

Ambient Temperature Requirements

5-40° C (40-104° F)

Measurement

Measurement 30-900 s

Dynamic Range

Standard: 0.13 ppm to 3000 ppm Extended Range (XR): 0.13 ppm to 4 wt%

NOW AVAILABLE WITH:

  1. ACCU-FLOW
  2. AUTOSAMPLER

Ordering Information

Instrument: 

Standard XRF Sample Cups - Bag of 100: 203688-01

Mylar thin-film: 200335-01

Calibration Standard: 



LEARN MORE HERE