Automatic Laboratory Turbidimeters the TU5200 is an essential part of your turbidity monitoring program. Let Hach Service help ensure that your instrument operates at peak performance to deliver the accurate, low-level measurements that you require for regulatory reporting and process management
Key advantages
- Ensures critical turbidity measurements are accurate and reliable
- Provides documentation to support regulatory compliance
- Detects and prevents issues that can degrade performance or reduce instrument life
- Loaner instruments ensure continuous operations during service
- Technical support hotline provides rapid and direct resolution to many issues
Essential Maintenance Tasks Performed
Routine:
- Perform visual inspection of housing for signs of damage or contamination • Examine and clean vial compartment
- Update firmware to latest version
- Calibrate with certified StablCal
- Perform hardware check to verify functionality of critical components and detect potential issues that may impact system performance
- Perform service report and performance certificate to support regulatory compliance
- Perform advanced troubleshooting to resolve any error conditions
Specifications
Measurement method
Nephelometry with the scattered light collected at a 90° angle to the incident light and 360° around the sample vial.
Primary compliance method
EPA approved Hach Method 102581
Dimensions (W x D x H)
41 x 28 x 12.5 cm (16 x 11 x 7.7 in.)
Weight
2.37 kg (5.23 lb)
Enclosure
IP20
Protection class
Instrument: III; Power supply: I
Laser
Class 2 laser product: Contains a non user-serviceable class 2 laser.
Accuracy
± 2% of reading plus 0.01 NTU from 0 to 40 NTU ± 10% of reading from 40 to 700 NTU based on Formazin primary standard at 25 °C (77 °F)
Ordering Information
Instrument: LPV442.99.03022