Sindie 7039 Sulfur Analyzer


RPV VALVES

Sindie 7039 complies with ASTM D7039 and ISO 20884 methods, and enables fast batch testing from 0.15 – 3000 ppm for sulfur fuel samples at petroleum pipeline terminals, refineries, and test laboratories. This unit is compact and fits on any lab bench with an easy-to-use and robust design requiring minimal maintenance. Sindie 7039 offers many advantages over competing technologies: It has exceptional signal-to-noise ratio, and does not require consumable gases or high-temperature operations.

Key advantages

  • LOD: 0.4 ppm at 300 s
  • Dynamic Range: 0.4 ppm - 3000 ppm
  • Easy to use - Intuitive touch screen - Just plug-in and measure - Measurement time: 30-900 s
  • Extremely low maintenance: no conversion gasses, heating elements, columns, or quartz tubing
  • 75 W air-cooled excitation tube
  • Fits on any lab bench

Applications

  • Total sulfur analysis from ultra low sulfur fuels to crudes
  • For use in refinery labs, pipeline terminals, additive plants, testing vans and inspection laboratories


TRUSTED PRECISION

Monochromatic Wavelength Dispersive X-ray Fluorescence (MWDXRF®) utilizes state-of-the-art focusing and monochromating optics to increase excitation intensity and dramatically improve signal-to-background over high power traditional WDXRF instruments. This enables significantly improved detection limits and precision, and a reduced sensitivity to matrix effects. A monochromatic and focused primary beam excites the sample and secondary characteristic fluorescence X-rays are emitted from the sample. A second monochromating optic selects the sulfur characteristic X-rays and directs these X-rays to the detector. MWDXRF is a direct measurement technique and does not require consumable gasses or sample conversion.

Specifications

Model

Sindie 7039 Gen 2

Test Method

ASTM D7039 and ISO 20884

Sample Cup Volume

10 mL

Ambient Temperature 5- Requirements

40° C (40-104° F)

Dynamic Range

Standard: 0.4 ppm - 3000 ppm

Extended Range (XR): 0.4 ppm - 10 wt%

Measurement

User selectable: 30-900 s

Calibration

8 calibration curves. Automatic and manual calibration functionality

Method Compliance

ASTM D4294, ISO 8754 & IP 336

Ordering Information

Instrument: 403101-01

Standard XRF Sample Cups - Bag of 100: 203688-01

Mylar thin-film: 200335-01

Calibration Standard: 204454-01



LEARN MORE HERE