Sindie 7039 complies with ASTM D7039 and ISO 20884 methods, and enables fast batch testing from 0.15 – 3000 ppm for sulfur fuel samples at petroleum pipeline terminals, refineries, and test laboratories. This unit is compact and fits on any lab bench with an easy-to-use and robust design requiring minimal maintenance. Sindie 7039 offers many advantages over competing technologies: It has exceptional signal-to-noise ratio, and does not require consumable gases or high-temperature operations.
Key advantages
- LOD: 0.4 ppm at 300 s
- Dynamic Range: 0.4 ppm - 3000 ppm
- Easy to use - Intuitive touch screen - Just plug-in and measure - Measurement time: 30-900 s
- Extremely low maintenance: no conversion gasses, heating elements, columns, or quartz tubing
- 75 W air-cooled excitation tube
- Fits on any lab bench
Applications
- Total sulfur analysis from ultra low sulfur fuels to crudes
- For use in refinery labs, pipeline terminals, additive plants, testing vans and inspection laboratories
TRUSTED PRECISION
Monochromatic Wavelength Dispersive X-ray Fluorescence (MWDXRF®) utilizes state-of-the-art focusing and monochromating optics to increase excitation intensity and dramatically improve signal-to-background over high power traditional WDXRF instruments. This enables significantly improved detection limits and precision, and a reduced sensitivity to matrix effects. A monochromatic and focused primary beam excites the sample and secondary characteristic fluorescence X-rays are emitted from the sample. A second monochromating optic selects the sulfur characteristic X-rays and directs these X-rays to the detector. MWDXRF is a direct measurement technique and does not require consumable gasses or sample conversion.
Specifications
Model
Sindie 7039 Gen 2
Test Method
ASTM D7039 and ISO 20884
Sample Cup Volume
10 mL
Ambient Temperature 5- Requirements
40° C (40-104° F)
Dynamic Range
Standard: 0.4 ppm - 3000 ppm
Extended Range (XR): 0.4 ppm - 10 wt%
Measurement
User selectable: 30-900 s
Calibration
8 calibration curves. Automatic and manual calibration functionality
Method Compliance
ASTM D4294, ISO 8754 & IP 336
Ordering Information
Instrument: 403101-01
Standard XRF Sample Cups - Bag of 100: 203688-01
Mylar thin-film: 200335-01
Calibration Standard: 204454-01